Table 2 Electrical features at implant and during follow-up.
Implant procedure | Follow-up | |||||
---|---|---|---|---|---|---|
Normal PCT group | Very high PCT group | P-value | Normal PCT group | Very high PCT group | P-value | |
(n = 390) | (n = 19) | (n = 390) | (n = 19) | |||
Pacing threshold, V/0.24 ms | 0.59 ± 0.32 | 0.90 ± 0.45 | < 0.001 | 0.51 ± 0.16 | 1.82 ± 0.64 | < 0.001 |
Pacing impedance, Ω | 858.7 ± 270.4 | 714.4 ± 199.2 | 0.026 | 613.87 ± 141.54 | 480.53 ± 67.37 | < 0.001 |
R wave amplitude, mV | 11.5 ± 38.3 | 9.8 ± 4.2 | 0.810 | 11.94 ± 4.08 | 9.85 ± 3.85 | 0.035 |
Multiple deployments (≥ 2) required, % | 16 (4.1%) | 0 (0.0%) | 0.368 | |||
Follow-up duration (days) | 354.55 ± 310.87 | 287.42 ± 215.82 | 0.680 |