Table 2 Electrical features at implant and during follow-up.

From: Retrospective evaluation of the predictive value of acute pacing capture threshold for long-term outcomes in Chinese patients with leadless pacemakers

 

Implant procedure

Follow-up

Normal PCT group

Very high PCT group

P-value

Normal PCT group

Very high PCT group

P-value

(n = 390)

(n = 19)

 

(n = 390)

(n = 19)

 

Pacing threshold, V/0.24 ms

0.59 ± 0.32

0.90 ± 0.45

< 0.001

0.51 ± 0.16

1.82 ± 0.64

< 0.001

Pacing impedance, Ω

858.7 ± 270.4

714.4 ± 199.2

0.026

613.87 ± 141.54

480.53 ± 67.37

< 0.001

R wave amplitude, mV

11.5 ± 38.3

9.8 ± 4.2

0.810

11.94 ± 4.08

9.85 ± 3.85

0.035

Multiple deployments (≥ 2) required, %

16 (4.1%)

0 (0.0%)

0.368

   

Follow-up duration (days)

   

354.55 ± 310.87

287.42 ± 215.82

0.680

  1. PCT indicates pacing capture threshold;