Table 1 Protocol and scanner settings.

From: Radiation dose optimisation in paediatric head CT using attenuation-based auto prescription

Age range

 

Age based protocol

Attenuation based protocol

All patients

Scanmode

Axial wide detector

Axial wide detector

Kernel and filter

Head

Head

Image reconstruction

DLIR high

DLIR high

0–1 years

Scan field of view

Ped head

Ped head

kV

100

100

mA (minimum/maximum)

50/420

50/420

dw, in cm

< 14.6

< 14.6

Noise index

2.8

3.2

Rotation time, in seconds

0.5

0.5

1–5 years

Scan field of view

Small head

Small head

kV

120

120

mA (minimum/maximum)

50/425

50/425

dw (minimum/maximum), in cm

14.6/15.8

14.6/15.8

Noise index

2.5

2.9

Rotation time, in seconds

0.35

0.35

5–10 years

Scan field of view

Small head

Small head

kV

120

120

mA (minimum/maximum)

50/405

50/405

dw (minimum/maximum), in cm

15.8/17.1

15.8/17.1

Noise index

2.6

3.4

Rotation time, in seconds

0.6

0.6

10–16 years

Scan field of view

Head

Head

kV

120

120

mA (minimum/maximum)

50/405

50/405

dw (minimum/ maximum), in cm

17.1/18.2

17.1/18.2

Noise index

2.6

3.4

Rotation time, in seconds

0.6

0.6

16 plus years

Scan field of view

Head

Head

kV

120

120

mA (minimum/maximum)

50/480

60/270

dw (minimum/maximum), in cm

18.2/25

18.2/25

Noise Index

2.6

3.4

Rotation time, in seconds

0.8

0.8

  1. DLIR deep learning image reconstruction, kV kilovoltage, mA milliampere, dw water equivalent diameter.