Table 5 Estimation results, actual result and error rates of A-VMD algorithm.

From: FMCW-based contactless heart rate monitoring

Experiment no.

Measurement distance (m)

Subject state

Breathing actual result

(times/min)

A-VMD measurement result (times/min)

A-VMD absolute error rate(%)

1

0.4

a A

15

15

0.00

2

0.4

b B

17

16

6.25

3

0.4

c A

14

14

0.00

4

0.4

a C

22

21

4.76

5

0.4

e C

30

29

3.45

6

0.6

d B

16

16

0.00

7

0.6

f B

18

17

5.88

8

0.6

g B

18

18

0.00

9

0.6

h B

17

17

0.00

10

0.6

i C

28

27

3.70

11

0.8

j A

14

14

0.00

12

0.8

b A

15

15

0.00

13

0.8

e B

17

17

0.00

14

0.8

d C

24

23

4.35

15

0.8

g C

31

30

3.33