Table 1 High-speed handpiece and three ways syringe output water contamination at different period of the semester.
Time point | Colony count (CFU/ml)a | T value | P value | Contamination rate (%) | ||
---|---|---|---|---|---|---|
High-speed handpiece | Three ways syringe | High-speed handpiece | Three ways syringe | |||
Beginning | 11,586\(\:\pm\:\)1715 | 5375\(\:\pm\:\)874 | 3.954 | 0.001 | 100.00 | 100.00 |
Middle | 2041\(\:\pm\:\)327*** | 1690\(\:\pm\:\)295** | 0.866 | 0.401 | 100.00 | 100.00 |
End | 1680\(\:\pm\:\)131*** | 1372\(\:\pm\:\)119** | 1.587 | 0.135 | 100.00 | 100.00 |