Key Points
-
QLF is a new device to detect very early demineralisation of enamel.
-
QLF can be used in vivo and in vitro.
-
All diagnostic methods suffer from some form of variability.
-
This study determined that, in most cases, QLF does not suffer from significant reliability issues in the analysis stage.
-
Further work on in vivo lesions is required.
-
Novices should be trained before analysing experimental data.
Abstract
Objective To assess the reliability of the analysis stage of quantitative light-induced fluorescence (QLF). The QLF analysis involves subjective input from the user and this study examines the influence of this on the reproducibility of the QLF data.
Method QLF images were taken of 20 human molar teeth that had been previously subjected to a demineralising solution (phosphoric acid 37%) to create artificial white spot lesions on their buccal surfaces. Following examination of the images, 16 were chosen to represent a range of lesion size and severity. Three copies were made of the images and each was allocated a different filename. 10 examiners in three centres were asked to analyse each of the 16 images on three occasions, with at least seven days between each attempt. Simple instructions describing the analysis procedure were supplied and examiners were asked to adhere to these directions. Examiners were asked to rate each of the 16 teeth on their first attempt both quantitatively (5 point scale) and qualitatively in terms of difficulty of analysis. Data reported were the ΔQ at 5% threshold for each tooth on each of three attempts.
Results Using ANOVA and paired t-tests to detect statistical differences, the three attempts of each examiner were used to determine intra-examiner reliability. Only one examiner (a novice at the technique) demonstrated differences between all three attempts and two demonstrated difference between one attempt. When the mean scores were compared to determine the inter-examiner reliability, only one examiner's results were statistically different when compared with two others.
Conclusion This study has demonstrated that the analysis stage of QLF is reliable between examiners and within multiple attempts by the same examiner, when analysing in vitro lesions. Novices at the technique should be trained before analysing experimental data.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
References
al-Khateeb S, ten Cate JM, Angmar-Mansson B et al. Quantification of formation and remineralization of artificial enamel lesions with a new portable fluorescence device. Adv Dent Res 1997; 11: 502–506.
Ando M, Hall AF, Eckert GJ, Schemehorn BR, Analoui M, Stookey GK Relative ability of laser fluorescence techniques to quantitate early mineral loss in vitro. Caries Res 1997; 31: 125–31.
Lagerweij M, vander Veen M, Ando M, Lukantsova L, Stookey G The validity and repeatability of three light-induced fluorescence systems: an in vitro study. Caries Res 1999; 33: 220–226.
Ross G Caries diagnosis with the DIAGNOdent laser: a user's product evaluation. Ont Dent 1999; 76: 21–24.
Fleiss JL The design and analysis of clinical experiments New York: John Wiley & Sons 1986.
Brunette DM Critical thinking: understanding and evaluating dental research London: Quintessence Pub. Co 1996.
Anderson P, Elliott JC Scanning X-ray microradiographic study of the formation of caries-like lesions in synthetic apatite aggregates. Caries Res 1985; 19: 403–406.
Author information
Authors and Affiliations
Corresponding author
Additional information
Refereed paper
Rights and permissions
About this article
Cite this article
Pretty, I., Hall, A., Smith, P. et al. The intra- and inter-examiner reliability of quantitative light-induced fluorescence (QLF) analyses. Br Dent J 193, 105–109 (2002). https://doi.org/10.1038/sj.bdj.4801496
Received:
Accepted:
Published:
Issue date:
DOI: https://doi.org/10.1038/sj.bdj.4801496
This article is cited by
-
Comparison of Quantitative light-induced fluorescence-digital (QLF-D) images and images of disclosed plaque for planimetric quantification of dental plaque in multibracket appliance patients
Scientific Reports (2020)
-
Quantitative light-induced fluorescence images and digital photographs - Reproducibility of manually marked demineralisations
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie (2017)
-
Assessment of a new infrared laser transillumination technology (808 nm) for the detection of occlusal caries—an in vitro study
Lasers in Medical Science (2015)
-
Demineralization adjacent to orthodontic brackets after application of conventional and self-etching primer systems
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie (2014)
-
Performance of a new fluorescence camera for detection of occlusal caries in vitro
Lasers in Medical Science (2013)