Abstract
Aims
To compare retinal nerve fibre layer (RNFL) measurements were carried out with two different versions of an optical coherence tomography device in patients with band atrophy (BA) of the optic nerve and in normal controls.
Methods
The RNFL of 36 eyes (18 with BA and 18 normals) was measured using an earlier version of an optical coherence tomography device (OCT-1). The measurements were repeated using a later version of the same equipment (OCT-3), and the two sets of measurements were compared.
Results
Using OCT-1, the peripapillary RNFL thickness (mean±SD, in μm) in eyes with BA measured 80.42±6.94, 99.81±14.00, 61.69±13.02, 101.70±12.54, and 57.36±16.52 corresponding to the total RNFL average, superior, temporal, inferior, and nasal quadrants, respectively. Using OCT-3, the corresponding measurements were 63.11±6.76, 81.22±13.34, 39.50±8.27, 86.72±15.16, and 45.05±8.03. Each of these measurements was significantly smaller with OCT-3 than with OCT-1. In normal eyes, RNFL average and temporal quadrant OCT-3 values were significantly smaller than OCT-1 values, but there was no significant difference in measurements from the superior, inferior, and nasal quadrant.
Conclusions
RNFL measurements were smaller with OCT-3 than with OCT-1 for almost all parameters in eyes with BA and in the global average and temporal quadrant measurements in normal eyes. Investigators should be aware of this fact when comparing old RNFL measurement with values obtained with later versions of the equipment.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
References
Chauhan DS, Marshall J . The interpretation of optical coherence tomography images of the retina. Invest Ophthalmol Vis Sci 1999; 40: 2332–2342.
Huang Y, Cideciyan AV, Papastergiou GI . Relation of optical coherence tomography to microanatomy in normal and rd chickens. Invest Ophthalmol Vis Sci 1998; 39: 2405–2416.
Toh CA, Narayan DG, Boppart SA, Hee MR, Fujimoto JG, Birngruber R et al. A comparison of retinal morphology viewed by light microscopy. Arch Ophthalmol 1997; 115: 1425–1428.
Jones AL, Sheen NJL, North RV, Morgan JE . The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layer. Br J Ophthalmol 2001; 85: 673–677.
Schuman JS, Hee MR, Puliafito CA, Wong C, Pedut-Kloizman T, Lin CP et al. Quantification of nerve fibre layer thickness in normal and glaucomatous eyes using optical coherence tomography. Arch Ophthalmol 1995; 113: 586–596.
Hoh ST, Greenfield DS, Mistlberger A, Liebmann JM, Ishikawa H, Ritch R . Optical coherence tomography and scanning laser polarimetry in normal, ocular hypertensive and glaucomatous eyes. Am J Ophthalmol 2000; 129: 129–135.
Parisi V, Manni G, Centofanti M, Gandolfi SA, Olzi D, Bucci MG . Correlation between optical coherence tomography, pattern electroretinogram, and visual evoked potentials in open-angle glaucoma patients. Ophthalmology 2001; 108: 905–912.
Furuichi M, Kashiwagi K, Furuichi Y, Tsukahara S . Comparison of the effectiveness of scanning laser polarimetry and optical coherence tomography for estimating optic nerve fibre layer thickness in patients with glaucoma. Ophthalmologica 2002; 216: 168–174.
Greaney MJ, Hoffman DC, Garway-Heath DF, Nakla M, Coleman AL, Caprioli J . Comparison of optic nerve imaging methods to distinguish normal eyes from those with glaucoma. Invest Ophthalmol Vis Sci 2002; 43: 140–145.
Monteiro MLR, Leal BC, Rosa AAM . Optical coherence tomography analysis of axonal loss in band atrophy of the optic nerve. Br J Ophthalmol 2004; 88: 896–899.
Kanamori A, Nakamura M, Matsui N, Nagai A, Nakanishi Y, Kusubara S et al. Optical coherence tomography detects characteristic retinal nerve fiber layer thickness corresponding to band atrophy of the optic discs. Ophthalmology 2004; 111: 2278–2283.
Blumenthal EZ, Williams JM, Weinreb RN, Girkin CA, Berry CC, Zangwill LM . Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography. Ophthalmology 2000; 107: 2278–2282.
Jones AL, Sheen NJL, North RV, Morgan JE . The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layer. Br J Ophthalmol 2001; 85: 673–677.
Villain MA, Greenfield DS . Peripapillary nerve fiber layer thickness measurement reproducibility using optical coherence tomography. Ophthalmic Surg Lasers Imaging 2003; 34: 33–37.
Jaffe GJ, Caprioli J . Optical coherence tomography to detect and manage retinal disease and glaucoma. Am J Ophthalmol 2004; 137: 156–169.
Bland JM, Altman DG . Statistical methods for assessing agreement between two methods of clinical measurements. Lancet 1986; 1: 307–310.
Bland JM, Altman DG . Measuring agreement in method of comparison studies. Stat Methods Med Res 1999; 8: 135–160.
Schuman JS, Pedut-Kloizman T, Hertzmark E, Hee MR, Wilkins JR, Coker JG et al. Reproducibility of nerve fiber layer thickness measurements using optical coherence tomography. Ophthalmology 1996; 103: 1889–1898.
Paunescu LA, Schuman JS, Price LL, Stark PC, Beaton S, Ishikawas H et al. Reproducibility of nerve fiber thickness, macular thickness, and optic nerve head measurements using Stratus OCT. Invest Ophthalmol Vis Sci 2004; 45: 1716–1724.
Kanamori A, Nakamura M, Escano MFT, Seya R, Maeda H, Negi A . Evaluation of the glaucomatous damage on retinal nerve fiber layer thickness measured by optical coherence tomography. Am J Ophthalmol 2003; 135: 513–520.
Bourne RAR, Medeiros FA, Bowd C, Jahanbakhsh D, Zangwill LM, Weinreb RN . Comparability of retinal nerve fiber layer thickness measurements of optical coherence tomography instruments. Invest Ophthalmol Vis Sci 2005; 46: 1280–1285.
Ishikawa H, Piette S, Liebmann JM, Ritch R . Detecting the inner and outer borders of the retinal nerve fiber layer using optical coherence tomography. Graefe's Arch Clin Exp Ophthalmol 2002; 240: 362–371.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Monteiro, M., Leal, B., Moura, F. et al. Comparison of retinal nerve fibre layer measurements using optical coherence tomography versions 1 and 3 in eyes with band atrophy of the optic nerve and normal controls. Eye 21, 16–22 (2007). https://doi.org/10.1038/sj.eye.6702182
Received:
Accepted:
Published:
Issue date:
DOI: https://doi.org/10.1038/sj.eye.6702182
Keywords
This article is cited by
-
Macular thickness measurements with frequency domain-OCT for quantification of axonal loss in chronic papilledema from pseudotumor cerebri syndrome
Eye (2014)
-
Correlation between multifocal pattern electroretinography and Fourier-domain OCT in eyes with temporal hemianopia from chiasmal compression
Graefe's Archive for Clinical and Experimental Ophthalmology (2013)
-
Correlation between macular and retinal nerve fibre layer Fourier-domain OCT measurements and visual field loss in chiasmal compression
Eye (2010)
-
Comparison of the GDx VCC scanning laser polarimeter and the stratus optical coherence tomograph in the detection of band atrophy of the optic nerve
Eye (2008)
-
Pattern electroretinograms for the detection of neural loss in patients with permanent temporal visual field defect from chiasmal compression
Documenta Ophthalmologica (2008)