To identify the source of unwanted glare images from a three-piece intraocular lens (IOL) implant following cataract surgery.
Method
The IOL and posterior capsule were examined under mydriatic and nonmydriatic conditions using direct focal illumination from a slit lamp biomicroscope. Direct focal illumination was undertaken with both a narrow beam (0.1 mm in width) and small spot (0.1 mm in diameter) to identify the points at which the glare images were stimulated. While observing the location of the beam with the slit lamp biomicroscope, the patient indicated when the glare images were stimulated.
Results
The nasal haptic insertion into the optic was identified as the source of temporal line images arising from lights such as headlamps from oncoming cars and street lamps. The adjacent edge of the IOL was also identified as the likely source of additional cob web-like light rays.
Conclusions
The haptic insertions in three-piece IOLs may, under certain conditions, interfere with light entering the pupil and produce extraneous images. Large mesopic pupils and decentred IOLs are conditions that increase the likelihood of unwanted glare images.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
References
Alpar JJ . Glare factors in different intraocular lenses: Comparative study in patients with dissimilar binocular implants, in Current Concepts in Cataract Surgery, Selected Proceedings of the Seventh Biennial Cataract Surgical Congress, Appleton-Century-Crofts/New York. 1982; 285–290.
Rosner M, Shari M, Blumenthal M . Optical aberrations from a well-centered intraocular lens implant. Am J Ophthalmol 1986; 101: 117–118.
Landry R . Unwanted optical effects caused by intraocular lens positioning holes. J Cataract Refract Surg 1987; 13: 421–423.
Burk PO . Investigation of unwanted images caused by intraocular lenses. J Cataract Refract Surg 1988; 21: 334–338.
Masket S, Geraghty E, Crandall AS, Davison JA, Johnson SH, Koch DD et al. Undesired light images associated with ovoid intraocular lenses. J Cataract Refract Surg 1993; 19: 690–694.
Holladay JT, Lang A, Portney V . Analysis of edge glare phenomenon in intraocular lens edge designs. J Cataract Refract Surg 1999; 25: 748–752.
Mamalis N . Complications of foldable intraocular lenses requiring explantation or secondary intervention-1998 survey. J Cataract Refract Surg 2000; 26: 766–772.
Kruger AJ, Schauersberger J, Abela C, Schild G, Amon M . Two year results: sharp versus rounded optic edges on silicone lenses. J Cataract Refract Surg 2000; 26: 566–570.
Davison JA . Positive and negative dysphotopia in patients with acrylic intraocular lenses. J Cataract Refract Surg 2000; 26: 11346–11355.
Masket S . Truncated edge design, dysphotopia, and inhibition of posterior capsule opacification. J Cataract Refract Surg 2000; 26: 145–147.
Farbowitz MA, Zabriskie NA, Crandall AS, Olson RJ, Miller M . Visual complaints associated with the AcrySof acrylic intraocular lens. J Cataract Refract Surg 2000; 26: 1339–1345.
Javitt J, Brauweiler H-P, Jacobi KW, Klemen U, Kohnen S, Quentin C-D et al. Cataract extraction with multifocal intraocular lens implantation: clinical, functional, and quality-of-life outcomes. J Cataract Refract Surg 2000; 26: 1356–1366.
Haring G, Dick HB, Krummenauer F, Weissmantel U, Kroncke W . Subjective photic phenomenon with refractive multifocal and monofocal intraocular lenses. J Cataract Refract Surg 2001; 27: 245–249.
Ellis MF . Sharp-edged intraocular lens design as a cause of permanent glare. J Cataract Refract Surg 2001; 27: 1061–1064.
Erie JC, Bandhauer MH, McLaren JW . Analysis of postoperative glare and intraocular lens design. J Cataract Refract Surg 2001; 27: 614–621.
Mamalis N, Spencer TS . Complications of foldable intraocular lenses requiring explantation or secondary intervention-2000 survey update. J Cataract Refract Surg 2000; 27: 1310–1317.
Serdarevic O . How to choose the right IOL for your cataract patient. Ophthalmol Times 2002; 21.
Author information
Authors and Affiliations
Corresponding author
Additional information
Financial/proprietary interest: The authors do not have any commercial or proprietary interest in any of the products referred to in this paper
Rights and permissions
About this article
Cite this article
Wolffe, M., Landry, R. & Alpar, J. Identification of the source of permanent glare from a three-piece IOL. Eye 21, 1078–1082 (2007). https://doi.org/10.1038/sj.eye.6702539
Received:
Accepted:
Published:
Issue date:
DOI: https://doi.org/10.1038/sj.eye.6702539