Abstract
Purpose To compare the intraocular pressure (IOP) measurements taken at the central cornea with those taken at the mid-peripheral cornea using the Tono-Pen tonometer.
Methods Forty-eight normal human subjects had their IOP measurements taken (i) with the Goldmann tonometer, (ii) with the Tono-Pen at the central cornea and (iii) with the Tono-Pen at the mid-peripheral cornea. The central and mid-peripheral corneal thickness at the site of IOP measurements were then determined by ultrasonic pachymetry.
Results The mean IOP with the Goldmann tonometer was 15.7 ± 3.1 mmHg. The mean Tono-Pen IOPs at the central and mid-peripheral cornea were 15.2 ± 3.2 mmHg and 15.7 ± 3.3 mmHg respectively. The mean mid-peripheral corneal thickness (578 µm) was sigificantly higher than the mean central thickness (538 µm) (p < 0.001).
Conclusion There was no significant difference between the IOP taken with the Tono-Pen and with the Goldmann tonometer. In addition, no clinically significant difference was observed between the IOP readings of central and mid-peripheral cornea measured by the Tono-Pen.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
References
Argus WA . Ocular hypertension and central corneal thickness. Ophthalmology 1995;102:1810–12.
Whitacre MM, Richard AS, Hassanein K . The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993;115:592–6.
Herndon LW, Choudhri SA, Cox T, Damji KF, Shields MB, Allingham R . Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes. Arch Ophthalmol 1997;115:1137–41.
Wolfs RCW, Klaver CCW, Vingerling JR, Grobbee DE, Hofman A, Dejong PTVM . Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam study. Am J Ophthalmol 1997;123:767–72.
Chatterjee A, Shah S, Bessant DA, Naroo SA, Doyle SJ . Reduction in intraocular pressure after excimer laser photorefractive keratectomy. Ophthalmology 1997;104:355–9.
Foster PJ, Bassanhu J, Alsbirk PH, Munkhbayar D, Uranchimeg D, Johnson GJ . Central corneal thickness and intraocular pressure in a Mongolian population. Ophthalmology 1998;105:969–73.
Ehlers N, Bramsen T, Sperling S . Applanation tonometry and central corneal thickness. Acta Ophthalmol (Copenh) 1975;53:34–43.
Kooner KS, Cooksey JC, Barron JB, Zimmerman TJ, Gupte RK, Wall JL . Tonometry comparison: Goldmann versus Tono-Pen. Ann Ophthalmol 1992;24:29–36.
Frenkel RE, Hong YJ, Shin DH . Comparison of the Tono-Pen to the Goldmann applanation tonometer. Arch Ophthalmol 1988;106:750–3.
Kao SF, Lichter PR, Bergstrom TJ, Rowe S, Musch DC . Clinical comparison of the Oculab Tono Pen to the Goldmann applanation tonometer. Ophthalmology 1987;94:1541–4.
Boothe WA . The Ton-Pen: a manometric and clinical study. Arch Ophthalmol 1988;106:1214–7.
Hessemer V, Rosier R, Jacobi KW . Comparison of intraocular pressure measurements with the Oculab Tono-Pen vs manometry in humans shortly after death. Am J Ophthalmol 1988;105:678–82.
Minckler DS, Baerveldt G, Heuer DK, Quillen-Thomas B, Walonker AF, Weiner J . Clinical evaluation of the Oculab Tono-Pen. Am J Ophthalmol 1987;104:168–73.
Khan JA, Davis M, Graham CE, Trank J, Whitacre MM . Comparison of Oculab Tono-Pen readings obtained from various corneal and scleral locations. Arch Ophthalmol 1991;109:1444–6.
Author information
Authors and Affiliations
Corresponding author
Additional information
The authors have no direct or indirect proprietary interests in the Tono-Pen tonometer
Rights and permissions
About this article
Cite this article
Mok, K., Wong, CL. & Lee, VH. Tono-Pen tonometer and corneal thickness. Eye 13, 35–37 (1999). https://doi.org/10.1038/eye.1999.7
Received:
Revised:
Accepted:
Issue date:
DOI: https://doi.org/10.1038/eye.1999.7
Keywords
This article is cited by
-
Real-world comparison between the Tonopen and Goldmann applanation tonometry in a university glaucoma clinic
International Ophthalmology (2021)
-
The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes
International Ophthalmology (2020)
-
Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry
International Ophthalmology (2011)
-
Intraocular pressure changes following Descemet’s stripping with endothelial keratoplasty
Graefe's Archive for Clinical and Experimental Ophthalmology (2010)
-
New TonoPen XL: Comparison with the Goldmann tonometer
Eye (2001)