Table 1 Summary of automated devices studied in measuring blood pressure with atrial fibrillation
From: Measurement of blood pressure in people with atrial fibrillation
Devices with some evidence for accuracy | Devices shown to be inaccurate |
|---|---|
Clinic or home measurement | Accutracker 1 [12] |
Systolic: Microlife Watch BPA 100 Plus [8] | Copal UA-251 [12] |
Systolic: Philips SureSigns VSi [6] | Microlife BP A6 [13] |
Systolic: Welch Allyn Vital Signs 300 [5] | Omron HEM 711AC [14] |
Systolic and diastolic: Tensoval duo control [7] | Omron HEM-750CP [15] |
Takeda UA-751 [12] | |
Ambulatory measurement | Welch Allyn 52000 [12] |
Systolic: A&D TM-2430 device [9]. | |
Diastolic: SpaceLabs 90207 [11] |