Table 1 Technical parameters for MR imaging pulse sequences.

From: The application of magnetic susceptibility separation for measuring cerebral oxygenation in preterm neonates

 

T1w

T2w

pcASL

SWI

Sequence

3D FSPGR

3D CUBE

Multi-shot 3D fast spin-echo

3D spoiled GRE flow-compensated

Acquisition plane

Coronal

Sagittal

Axial

Axial

Phase-encoding direction

Left-Right

Posterior-Anterior

Posterior-Anterior

Left-Right

TR (ms)

7.74

2300

4680

30.9

TE (ms)

2.97

66.29

10.55

5 echoes; first echo: 5; echo spacing: 5.24

Flip angle

12°

90°

111°

20°

FOV (cm)

20

20

24

25

Acquisition matrix

512 × 512

256 × 256

128 × 1 28

256 × 256

In-plane resolution (mm)

0.39 × 0.39

0.78 ×0.78

1.875 × 1.875

0.977 × 0.977

Slice thickness (mm)

1

1

4

2, reconstructed to 1 with zero filling (ZIP2)

Number of slices

126

106

50

92

Additional parameters

n/a

n/a

1450 ms label period; 2025 ms pulse label; 24 control-label pairs

n/a

Scan duration

4 min 39 s

5 min 1 s

5 min 26 s

5 min 29 s

  1. T1w = T1-weighted, T2w = T2-weighted, pcASL pseudo-continuous arterial spin labeling, SWI susceptibility weighted imaging, FSPG fast spoiled gradient echo, CUBE General Electric name of sequence, not an acronym, GRE gradient echo, ZIP2 through-plane zero filling interpolation.