Table 1 Technical parameters for WBLDCT exams in MM
Number of detector rows | 16 or more |
Scan coverage | Cranial vault to proximal tibial metaphysis (include humeri in the field of view) |
Tube voltage(kV)/time-current product (mAs) | 120/50–70a |
Collimation | 0.5–1.5 mm |
Reconstruction convolution kernel | Sharp, high-frequency (bone) and smooth (soft tissue). Alternatively, one middle-frequency kernel for all images |
Iterative reconstruction algorithms | Yes (to reduce image noise and streak artifacts) |
Thickness/increment of axial slices | 2/1 mm or 3/1.5 mm |
Multiplanar Reconstructions (MPRs) | Yes (sagittal, coronal and parallel to long axis of proximal limbs). |