Table 2 Orthoptists and parent/carer’s iSight Test Pro measurements compared to standard clinical care measurements.
Difference between iSight test pro and clinical tests | Orthoptist iSight test pro measure N (%) | Parent/carer iSight test pro measure N (%) |
|---|---|---|
Within 1 line | 53 (63.1) | 39 (46.4) |
1–2 lines | 23 (27.4) | 21 (25.0) |
2–3 lines | 6 (7.1) | 12 (14.3) |
3–4 lines | 0 (0.0) | 4 (4.8) |
4–5 lines | 1 (1.2) | 5 (6.0) |
5–6 lines | 1 (1.2) | 1 (1.2) |
Greater than 6 lines | 0 (0.0) | 2 (2.4) |
Median difference to clinical standard (25th–75th percentile), logMAR | 0.060 (0.020–0.120) | 0.100 (0.045–0.200) |
Range of differences to clinical standard, logMAR | 0–0.500 | 0–1.280 |