Table 1 Predictors of the between-cuff systolic and diastolic blood pressure discrepancies in 31 obese men

From: In search of the optimal cuff for blood pressure measurement in people with severe obesity

Predictors

Coefficient

Standard error

T

p value

Systolic blood pressure discrepancy

  Age

−0.045

0.05

−0.90

n.s.

  Angle difference

−0.613

0.21

−2.97

0.004

  Systolic blood pressure

−0.131

0.04

−2.97

0.004

  Arm length

−1.113

0.27

−4.17

0.0001

  Mid-arm circumference

−0.283

0.32

−0.89

n.s.

  Skinfold thickness

−2.163

0.84

−2.57

0.012

  Base angle

1.436

0.58

2.48

0.015

Diastolic blood pressure discrepancy

  Age

−0.043

0.04

−0.99

n.s.

  Angle difference

−0.662

0.19

−3.53

0.0007

  Diastolic blood pressure

−0.141

0.06

−2.22

0.029

  Arm length

−0.380

0.24

−1.59

n.s.

  Mid-arm circumference

−0.196

0.29

−0.68

n.s.

  Skinfold thickness

−0.013

0.76

−0.02

n.s.

  Base angle

1.013

0.52

1.95

0.054

  1. Systolic blood pressure: Multiple R = 0.56; squared multiple R = 0.33.
  2. Diastolic blood pressure: Multiple R = 0.48; squared multiple R = 0.23.