Table 2 Participant cuff and invasive blood pressure measurements (n = 795)
From: Accuracy of cuff blood pressure and systolic blood pressure amplification
Cuff BP (mmHg) | |
SBP | 130 ± 18 |
DBP | 76 ± 11 |
PP | 54 ± 14 |
MAP | 94 ± 12 |
Heart rate (bpm) | 67 ± 12 |
Invasive brachial BP (mmHg) | |
SBP | 138 ± 22 |
DBP | 69 ± 10 |
PP | 69 ± 19 |
MAP | 92 ± 12 |
Invasive aortic BP (mmHg) | |
SBP | 130 ± 21 |
DBP | 70 ± 10 |
MAP | 90 ± 12 |
Invasive BP amplification (mmHg)a | |
SBP | 7.3 ± 9.1 |
DBP | −0.8 ± 4.4 |
PP | 8.1 ± 8.6 |
Cuff SBP corrected with SBP amplification (mmHg)b | 137 ± 21 |
Cuff BP differences (mmHg)c | |
Cuff SBP – invasive brachial SBP | −7.6 ± 11.9 |
Cuff DBP – invasive brachial DBP | 7.4 ± 8.3 |
Cuff PP – invasive brachial PP | −15.0 ± 12.6 |