Table 4 OECTs for other small molecules sensing

From: Biomolecule sensors based on organic electrochemical transistors

Analyte

Gate

Channel material

Electrolyte

Current sensitivity

Detection limit

Detection range

Ref.

Cortisol

Ag/AgCl

PEDOT:PSS

Artificial sweat

2.68 μA/decade

-

0.1 nM–0.5 mM

115

UA

Pt/Nafion/Graphene

/PANI/UOx-GO

PEDOT:PSS

PBS

-

30 nM

30 nM–0.3 mM

76

Urea

PEDOT:PSS

PEDOT:PSS

PBS

-

1 μM

1 μM–10 mM

123

AA

NOCC

PEDOT:PSS

PBS

--

0.26 μM-

0.3 μM–1.4 mM

5 μM–1 mM

87

Cortisol

Ag/AgCl

PEDOT:PSS

PBS

-

0.0088 fg/mL

1 fg/mL–1 μg/mL

116

Cortisol

Au

PEDOT:PSS

PBS

50 μA/decade

100 pM

1 nM–1 μM

114

UA

Hot-glued textile

PEDOT:PSS

Simulated wound exudate

0.46 NCR/decade

23 μM

100 μM–1 mM

122

UA

MIP/PEDOT/

carbon fiber

PEDOT/rGO/

cotton fiber

Gel

100 mA/decade

-

1 nM–500 μM

112

AA

PEDOT:PSS

PEDOT:PSS

PBS

-

-

10 μM–1200 μM

109

AA

PEDOT:PSS

PEDOT:PSS

PBS

61.2 μA/decade

80.1 μM

0.01 μM–0.1 mM

120