Extended Data Fig. 6: Reader antenna positioning.
From: Wirelessly operated bioelectronic sutures for the monitoring of deep surgical wounds

a, Illustration of the steps to position the reader antenna. b, Contour plot of the received harmonic signal power when the position of the antenna is scanned within a 40 mm × 40 mm area. c, Measured backscattering signal for the antenna positions in (b). Yellow shading indicates the 10 mm × 10 mm area with highest signal amplitude. d, Resonant frequency and received power as a function of the orientation angle of the antenna. Blue shading denotes the frequency uncertainty due to decrease in the signal-to-noise ratio. e, Harmonic backscattering spectra for varying orientation angles.