Table 2 Validity (Part A).

From: Multi-Planar Cervical Motion Dataset: IMU Measurements and Goniometer

  

Repetition 1

Repetition 2

Repetition 3

Mean x̄ ± SD

Sagittal

Flexion

r = 0.78

r = 0.78

r = 0.91

r = 0.82 ± 0.075

Extension

r = 0.88

r = 0.81

r = 0.81

r = 0.83 ± 0.04

Frontal

Side flexion left

r = 0.68

r = 0.64

r = 0.41

r = 0.58 ± 0.14

Side flexion right

r = 0.7

r = 0.74

r = 0.37

r = 0.61 ± 0.20

Horizontal

Rotation left

r = 0.30

r = 0.27

r = 0.18

r = 0.28 ± 0.06

Rotation right

r = 0.53

r = 0.50

r = 0.34

r = 0.46 ± 10

  1. Validity of cervical range of motion for each cervical movement in accordance with the description in Part A, i.e., both Flexion/Extension (Sagittal/Pitch), Side flexion Left/ Side flexion Right (Frontal/Roll) and Rotation Left/Rotation Right (Horizontal/Yow). Table indicates the Pearson r score for the Range of motion between the Universal Goniometer and the wearable internal measurement sensor for each cervical movement. Validation is provided both divided by repetition and for all repetitions.