Table 3 Accuracy for detecting refractive amblyopia risk factors.

From: Pediatric vision screening using the plusoptiX A12C photoscreener in Chinese preschool children aged 3 to 4 years

 

Age (month)

Myopia (D)

Hyper (D)

Astig (D)

Aniso (D)

Sensitivity (%)

Specificity (%)

PPV (%)

NPV (%)

Manufacturer (sensitivity)

30–50

≤−1.00

≥+1.00

≥1.00

≥1.00

100.00

49.57

7.45

100.00

Matta/Silbert

36–72

≤−1.00

≥+1.25

≥1.00

≥1.25

100.00

57.68

8.75

100.00

AAPOS 2013

31–48

<−3.00

>+4.00

>2.00

>2.00

92.86

94.49

40.63

99.69

ABCD 2012

8–72

≤−2.25

≥+2.50

≥2.25

≥1.00

100.00

90.72

30.43

100.00

Manufacturer (specificity)

36–72

≤−1.50

≥+2.50

≥1.50

≥1.00

100.00

82.03

18.42

100.00