Table 1 Optimised conditions for LA-SP-ICP-MS and quantitative analysis.

From: Imaging gold nanoparticles in mouse liver by laser ablation inductively coupled plasma mass spectrometry

LA parameters

LA-SP-ICP-MS values

Quantitative analysis values

ICP-MS parameters

LA-SP-ICP-MS values

Quantitative analysis values

Laser wavelength (nm)

213

213

RF power (W)

1400

1400

Laser energy (%)

5

10

Sampling depth

150

150

Laser frequency (Hz)

1

20

Signal acquisition mode

Time-resolved analysis

Time-resolved analysis

Ablation spot size (μm)

200

100

Dwell time (ms)

1

10

Carrying gas (He) (l/min)

0.7

0.7

Carrying gas (Ar) (l/min)

0.7

0.7

Scan rate (μm/s)

69

20

Isotopes

197Au

197Au