Table 1 Information from manufacturers kit manual and included in the kits.

From: Comparison of commercial ELISA assays for quantification of corticosterone in serum

 

DRG-5186

DRG-4164

Enzo

Arbor Assays

Standard series:

15–2250 ng/ml.

1.7–83 ng/ml

0.032–20 ng/ml.

0.078–10 ng/ml

Analytical sensitivity:

4.1 ng/ml

0.5647 ng/ml

0.02699 ng/ml

0.0186 ng/ml

Amount of sample (for dilution):

20 µl

10 µl

10 µl

5 µl

Sample dilution:

1:1

1:10

1:40

1:100

Release from CBG:

n.s.

n.s.

YES

YES

Internal quality control

Yes*

Yes

No

No

Cross reactivity (%):

 Progesterone

0.7

7.4

1.7

0.24

 Aldosterone

0.2

n.a.

0.18

0.62

 Cortisol

0.3

0.3

0.046

0.38

 11-Deoxycorticosterone

2.4

3.4

28.6

12.3

 Dehydroepiandrosterone

n.d.

n.a.

n.a.

n.a.

 Estriol

n.d.

n.a.

n.a.

n.a.

 Estradiol

n.d.

n.a.

n.a.

<0.08

 17-hydroxyprogesterone

n.d.

n.a.

n.a.

n.a.

 18-Hydroxydeoxycorticosterone

n.d.

n.a.

n.a.

n.a.

 Testosterone

n.d.

n.a.

0.13

n.a.

 Pregnolone Prognenolone

n.d.

0.3

<0.03

n.a.

 11-Dehydrocorticosterone

n.a.

1.6

n.a.

n.a.

 Tetrahydrocorticosterone

n.a.

n.a.

0.28

0.76

 β-Estradiol

n.a.

n.a.

<0.03

n.a.

 Cortisone

n.a.

n.a.

<0.03

<0.08

 11-dehydrocorticosterone acetate

n.a.

n.a.

<0.03

n.a.

 Dexamethasone

n.a.

n.a.

n.a.

0.12

 Corticosterone-21-Hemisuccinate

n.a.

n.a.

n.a.

<0.1

 Other steroids

n.a.

<0.1

n.a.

n.a.

  1. *Recommended quality control kit (DRG CTL-5262). CBG -Corticosteroid-binding globulin; n.s. not specified in the manual; n.d. cross reactivity not detected for these substances; n.a. the cross reactivity is not assessed for these substances.