Table 2 Summary of Participant Accuracy Over Test Blocks.

From: Neuroprosthetic-enabled control of graded arm muscle contraction in a paralyzed human

Description

Block #

Rest Angle (36°)

Angle A

Angle B

Angle C

Max Angle (128°)

Total Correct

Accuracy

Test 1

6

12/12

3/3 (60°)

0/3 (82°)

1/3 (105°)

0/3

16/24

66.7%

Test 2

7

12/12

3/3 (60°)

0/3 (82°)

0/3 (105°)

1/3

16/24

66.7%

Test 3

8

12/12

3/3 (60°)

0/3 (82°)

0/3 (105°)

1/3

16/24

66.7%

Test 4

9

12/12

3/3 (60°)

2/3 (82°)

3/3 (105°)

1/3

21/24

87.5%

Generalization Test 1

10

12/12

3/3 (57°)

1/3 (76°)

3/3 (111°)

3/3

22/24

91.7%

Generalization Test 2

11

12/12

3/3 (71°)

2/3 (90°)

2/3 (114°)

2/3

21/24

87.5%

Test pre-recalibration

6–8

36/36

9/9

0/12

1/12

2/12

48/72

66.7%

Test post-recalibration

9–11

36/36

9/9

5/9

8/9

6/9

64/72

88.9%

Generalization Blocks

10–11

24/24

6/6

3/6

5/6

5/6

43/48

89.6%

  1. Stimulation parameters were recalibrated between the third and fourth test block. Angles A, B, and C were at 60°, 82°, and 105° respectively for blocks 6–9 and were moved to angles not previously attempted—57°,76°, and 111° for the first generalization test block (Block 10) and 71°, 90°, and 114° for the second generalization test block (Block 11).