Table 1 Precision of the proposed assay.
From: Development of an Affimer-antibody combined immunological diagnosis kit for glypican-3
Sample concentration (ng/mL) | Intra-assay precision (n = 20) | Inter-assay precision (n = 10 * 3) | ||
|---|---|---|---|---|
Mean ± SD (ng/mL) | CV (%) | Mean ± SD (ng/mL) | CV (%) | |
1 | 0.92 ± 0.07 | 7.60 | 0.89 ± 0.08 | 8.98 |
5 | 4.29 ± 0.26 | 6.06 | 5.01 ± 0.40 | 7.98 |
50 | 44.07 ± 2.80 | 6.35 | 46.89 ± 3.35 | 7.14 |