Table 1 Results for testing of physical properties of test sealers in different environmental conditions ± SD.

From: Influence of environment on testing of hydraulic sealers

Material

Media

Test

Flow

Film Thickness

Radiopacity

Setting Time

mm

μm

mm Al

mins

AH Plus

as specified

24 ± 2

15 ± 5

16 ± 1

688

HBSS

/

/

/

433

DMEM

/

/

/

485

MTA Fillapex

as specified

30 ± 1

19 ± 4

4 ± 1

unset

HBSS

/

/

/

unset

DMEM

/

/

/

unset

BioRoot RCS

as specified

25 ± 1

46 ± 7

7 ± 1

235

HBSS

/

 /

/

667

DMEM

/

 /

/

650

Endoseal

as specified

25 ± 1

35 ± 5

12 ± 1

435

HBSS

/

/

/

662

DMEM

/

/

/

645

ISO Standard

6876

6876

6876

6876

value

>17

<50

<3

/