Table 1 Results for testing of physical properties of test sealers in different environmental conditions ± SD.
From: Influence of environment on testing of hydraulic sealers
Material | Media | Test | |||
|---|---|---|---|---|---|
Flow | Film Thickness | Radiopacity | Setting Time | ||
mm | μm | mm Al | mins | ||
AH Plus | as specified | 24 ± 2 | 15 ± 5 | 16 ± 1 | 688 |
HBSS | / | / | / | 433 | |
DMEM | / | / | / | 485 | |
MTA Fillapex | as specified | 30 ± 1 | 19 ± 4 | 4 ± 1 | unset |
HBSS | / | / | / | unset | |
DMEM | / | / | / | unset | |
BioRoot RCS | as specified | 25 ± 1 | 46 ± 7 | 7 ± 1 | 235 |
HBSS | / | / | / | 667 | |
DMEM | / | / | / | 650 | |
Endoseal | as specified | 25 ± 1 | 35 ± 5 | 12 ± 1 | 435 |
HBSS | / | / | / | 662 | |
DMEM | / | / | / | 645 | |
ISO Standard | 6876 | 6876 | 6876 | 6876 | |
value | >17 | <50 | <3 | / | |