Table 2 Accuracy of pedicle screw placement among three assisted technologies.
Screw Position* | Robot-PLIF (n [%]) | Template-PLIF (n [%]) | CT-based PLIF (n [%]) | P value | ||
|---|---|---|---|---|---|---|
RP vs TP | RP vs CP | TP vs CP | ||||
A | 924 (91.3) | 720 (81.3) | 1073(84.1) | <0.001 | <0.001 | 0.062 |
B | 48 (4.7) | 83 (9.4) | 114 (8.9) | <0.001 | <0.001 | 0.730 |
A + B | 972 (96.0) | 803(90.6) | 1187(93.0) | <0.001 | 0.019 | 0.043 |
C | 28 (2.8) | 33 (3.7) | 56 (4.4) | 0.238 | 0.040 | 0.445 |
D | 6 (0.6) | 14 (1.6) | 11 (0.9) | 0.036 | 0.456 | 0.125 |
E | 0 (0.0) | 6 (0.7) | 2 (0.2) | 0.009 | 0.208 | 0.050 |
R | 6 (0.6) | 30 (3.4) | 20 (1.6) | <0.001 | 0.029 | 0.006 |
Total | 1012 | 886 | 1276 | |||