Table 1 CTDI100 of different slice thicknesses measured by the 10-cm ionization chamber and the nMAG gel dosimeter at 120 kVp.

From: Dose and slice thickness evaluation with nMAG gel dosimeters in computed tomography

CTDI (mGy)

2 mm

5 mm

10 mm

14.4 mm

38.4 mm

CTDI100,c

 Ionization chamber

35.80

33.13

33.62

42.95

39.86

 nMAG gel

33.92

31.53

32.20

44.47

41.97

 Difference

−5.25%

−4.83%

−4.22%

3.54%

5.29%

CTDI100,p

 Ionization chamber

37.00

34.39

34.70

44.96

43.33

 nMAG gel

37.47

34.79

35.33

44.58

42.81

 Difference

1.27%

1.16%

1.82%

−0.85%

−1.20%

CTDI100,w

 Ionization chamber

36.60

33.97

34.34

44.29

42.18

 nMAG gel

36.29

33.70

34.29

44.54

42.53

 Difference

−0.85%

−0.79%

−0.15%

0.56%

0.83%