Table 2 CTDI100 of different slice thicknesses measured by the 10-cm ionization chamber and the nMAG gel dosimeter at 140 kVp.

From: Dose and slice thickness evaluation with nMAG gel dosimeters in computed tomography

CTDI (mGy)

2 mm

5 mm

10 mm

14.4 mm

38.4 mm

CTDI100,c

 Ionization chamber

50.42

46.30

46.86

60.01

55.00

 nMAG gel

47.86

44.84

45.45

60.74

57.47

 Difference

−5.08%

−3.15%

−3.01%

1.22%

4.49%

CTDI100,p

 Ionization chamber

51.57

47.57

47.54

62.53

58.73

 nMAG gel

52.21

48.43

48.06

62.35

58.11

 Difference

1.24%

1.81%

1.09%

−0.29%

−1.06%

CTDI100,w

 Ionization chamber

51.18

47.14

47.31

61.69

57.49

 nMAG gel

50.76

46.57

47.19

61.81

57.90

 Difference

−0.82%

−1.21%

−0.25%

0.19%

0.71%