Table 1 Fixed value stability test results table.

From: A Study on Micropipetting Detection Technology of Automatic Enzyme Immunoassay Analyzer

Actual value/mg

1

2

3

4

5

6

7

8

Average value/mg

CV/%

error/%

9.98

10.99

11.18

11.05

11.21

11.18

11.15

11.11

11.13

11.12

1.12

9.94

50.2

54.44

54.81

54.57

54.97

54.82

54.75

54.67

55.03

54.76

0.74

8.77

99.82

108.8

109.14

108.65

109.46

109.17

109.03

108.86

109.57

109.09

0.37

8.43