Figure 1
From: Hematoxylin and eosin staining of intact tissues via delipidation and ultrasound

Effectiveness of iHE. (A) Device configuration of iHE. (B) Effect of ultrasound on enhanced rinsing: the x-axis represents the rinsing time, and the y-axis represents the absorbance of the rinsing solution. The p value between the ultrasound and control groups showed a sharp decrease over time (Student’s t test, n = 5). After rinsing for 20 min, the difference between the ultrasound and control groups was significant (p < 0.001). Comparison of the two groups was performed by unpaired t-test (two tailed). All data are presented as the mean ± standard error of the mean. *p < 0.05, **p < 0.01, ***p < 0.001. (C) Enhancing the staining effect of DCM delipidation and ultrasound. C57BL/6 mouse brains were stained with hematoxylin and then were cut along the maximum sagittal plane. I With DCM delipidation but without ultrasound; staining for 6.5 h. II With DCM delipidation and with ultrasound; staining for 6 h. III Without DCM delipidation and without ultrasound; staining for 6.5 h. IV Without DCM delipidation but with ultrasound; staining for 6 h. All staining was performed at 50 °C.