Table 6 Comparisons of the vascular density, perfusion density, and foveal avascular zone metrics between two consecutive macular angiography scans showing different signal strength in the same subjects (n = 17).
1st measurements Signal strength = 8 | 2nd measurements Signal strength = 9 | p-value | |
|---|---|---|---|
Vessel density | |||
Total area (mean ± SD, mm−1) | 15.31 ± 2.13 | 16.64 ± 1.29 | < 0.001 |
Center (mean ± SD, mm−1) | 6.87 ± 3.05 | 7.91 ± 2.32 | 0.031 |
Inner - average (mean ± SD, mm−1) | 14.99 ± 2.63 | 16.31 ± 2.22 | 0.010 |
Outer - average (mean ± SD, mm−1) | 15.72 ± 2.09 | 17.06 ± 1.20 | < 0.001 |
Perfusion density | |||
Total area (mean ± SD) | 0.380 ± 0.054 | 0.414 ± 0.033 | 0.001 |
Center (mean ± SD) | 0.149 ± 0.072 | 0.173 ± 0.054 | 0.025 |
Inner - average (mean ± SD) | 0.349 ± 0.069 | 0.382 ± 0.058 | 0.012 |
Outer - average (mean ± SD) | 0.380 ± 0.054 | 0.414 ± 0.033 | < 0.001 |
Foveal avascular area | |||
Area (mean ± SD, mm2) | 0.18 ± 0.09 | 0.22 ± 0.10 | 0.042 |
Perimeter (mean ± SD, mm) | 1.83 ± 0.42 | 1.91 ± 0.41 | 0.118 |
Circularity (mean ± SD) | 0.67 ± 0.09 | 0.72 ± 0.08 | 0.046 |