Table 1 Summary of the parameters of the mounting media tested and their performance in 3D-SIM of thick samples and SMLM.

From: Aqueous mounting media increasing tissue translucence improve image quality in Structured Illumination Microscopy of thick biological specimen

Mounting medium for SIM:

Refractive Index @589 nm, 25 °C

Viscositya

3D-SIM of thick samples

SMLM Alexa 647

Hoechst 33258

Alexa 488

Alexa 555

Cy3

Alexa 594

Vectashield

1.4477

− − − −

+

+

+

+

+

+++

S65VE

1.4580

− − −

+++

+++c*

++

+++e*

+++

+++f*

Sorb70VE

1.4585

− − −

++++*

+++

++*

+++e

+++*

+++*

F80VE

1.4706

− − − − −b

+++

++

++

+++e

+++

++

H71VE

1.5037

− − − −

++++

+d

+++d

++

+++d

+

  1. Vectashield H-1000 was used in all experiments. Note a very low viscosity of intermediate solutions (Supplementary Table 2). Notes: aEstimates of the viscosity of embedding media are provided for respective temperatures of embedding. See Supplementary Table 1 for more details. Viscosity scales proportionally with the number of “−” signs. bHigh viscosity of fructose solutions can be decreased by an addition of urea41. This, however, may lead to DNA denaturation and impairment to Hoechst 33258 signal intensity. cHigh signal to noise ratio with faster bleaching. dSignal to noise ratio is low unless higher Vectashield vol/vol concentration in clearing media is used (Supplementary Fig. S10). eVery high signal to noise ratio. fElevating the Vectashield concentration to 40% vol/vol improves Alexa 647 photoncount (see Supplementary Fig. S14). *Recommended medium for the fluorophore. Due to fluorescence signal instability in the presence of Vectashield Alexa 647 is not recommended in SIM.