Table 1 Analytical recovery of P4 from urine in the 2 automated assays. SD; standard deviation. Architect: Abbott Architect progesterone chemiluminescent microparticle immunoassay, Cobas: Roche Cobas electrochemiluminescence immunoassay, P4: progesterone.
P4 added, nmol/L | % recovery, mean (SD) | |
|---|---|---|
Cobas | Architect | |
Male | ||
12.5 | 170 (41.7) | 113 (13.3) |
6.25 (x2 dilution) | 160 (30.0) | 106 (16.4) |
3.13 (x4 dilution) | 153 (26.6) | 109 (16.7) |
1.56 (x8 dilution) | 143 (22.0) | 112 (67.0) |