Figure 6 | Scientific Reports

Figure 6

From: MOMSense: Metal-Oxide-Metal Elementary Glucose Sensor

Figure 6

(a) A chart illustrating the measured current across MOMSense devices over time while 1 V is applied across the electrode and the glucose-containing sample. The results are shown for concentrations of 2.2 mM, 3.9 mM, 5.6 mM, 7.8 mM, 10.0 mM and 12.2 mM. These concentrations represent extremely low, low, medium, high and extremely high blood glucose levels for an adult human. As shown, the detected currents for each sample increase over time according to glucose concentration. Each concentration sample is tested on identical fresh devices. (b) A chart illustrating current data obtained with respect to Fig. 6(a) at a set time point of 18 s as a function of glucose concentration. It is clear that the response of the MOMSense devices saturates at a glucose concentration of 10.0 mM (180 mg/dl).

Back to article page