Table 1 Ventilatory parameters during the study protocol.

From: Intravenous administration of normal saline may be misinterpreted as a change of end-expiratory lung volume when using electrical impedance tomography

Parameter

Baseline values (Phase 1)

End values (Phase 6)

p Wilcoxon test

Expiratory tidal volume, mL

469 (426–494)

473 (424–498)

0.91

Respiratory rate*, b/min

20 (20–22)

20 (20–22)

Inspiratory to expiratory time ratio*

0.5

0.5

Fraction of inspired oxygen*, %

25 (25–30)

25 (25–30)

End-tidal carbon dioxide concentration, mmHg

43 (40–44)

42 (40–44)

0.12

Peripheral capillary oxygen saturation, %

98 (97–100)

98 (96–100)

0.91

Minute ventilation, L/min

9.2 (8.9–9.5)

9.3 (8.9–9.5)

0.47

Peak airway pressure, cmH2O

23 (21–26)

25 (24–26)

0.07

Positive end-expiratory pressure*, cmH2O

5

5

Compliance, ml/cmH2O

32 (31–42)

31 (30–39)

0.07

  1. Data are presented as median and interquartile range (Q1–Q3).
  2. *Preset parameter.