Table 2 Summary of dose falloff results in CyberKnife IRIS- and MLC-based plans.
Number of Plans | R50% | R10% | Ratio(MLC/IRIS) | P-value (Wilcoxon’s signed-rank) | |||||
|---|---|---|---|---|---|---|---|---|---|
IRIS | MLC | IRIS | MLC | R50% | R10% | R50% | R10% | ||
All cases | 15 | 5.52 ± 1.41 (5.74) | 4.64 ± 1.20 (4.56) | 54.00 ± 25.60 (45.74) | 54.99 ± 28.92 (45.56) | 0.84 | 1.02 | 0.006 | 0.865 |
Target volume ≤1 cc | 3 | 5.65 ± 1.17 (6.40) | 5.95 ± 1.17 (5.86) | 71.94 ± 18.97 (80.07) | 95.02 ± 8.89 (98.13) | 1.05 | 1.32 | 0.593 | 0.109 |
1 cc < Target volume ≤10 cc | 7 | 5.40 ± 1.83 (4.95) | 4.26 ± 1.18 (4.01) | 57.85 ± 29.30 (48.38) | 54.37 ± 24.94 (45.56) | 0.79 | 0.94 | 0.028 | 0.612 |
Target volume >10 cc | 5 | 5.61 ± 0.63 (5.74) | 4.39 ± 0.49 (4.20) | 37.84 ± 8.54 (37.15) | 31.84 ± 9.62 (32.06) | 0.78 | 0.84 | 0.043 | 0.043 |