Table 2 Accuracy of the IR-UWB radar in detecting PeAF.

From: Preclinical Evaluation of a Noncontact Simultaneous Monitoring Method for Respiration and Carotid Pulsation Using Impulse-Radio Ultra-Wideband Radar

 

ECG

PeAF (N = 14)

NSR (N = 19)

Radar

PeAF (N = 17)

14

3

Max∆frq 0.57 Hz

NSR (N = 16)

0

16

  1. Sensitivity = 1.00, specificity = 0.84, CCR = 0.909.
  2. Cohen’s Kappa = 0.819 (95% CI 0.627–1.000).