Table 3 One-way-ANOVA and pairwise comparison analysis of objective noise measurement among different sequences.
LD-MBIR | LD-ASIR | RD-ASIR | One-way-Anova | Pairwise comparison analysis | ||||
|---|---|---|---|---|---|---|---|---|
F-value | P-value | P-value (LD-MBIR and LD-ASIR) | P-value (LD-MBIR and RD-ASIR) | P-value (LD-ASIR and RD-ASIR) | ||||
left ventricles | 18.80 ± 2.99 | 72.25 ± 13.95 | 16.09 ± 2.55 | 441.18 | <0.001 | <0.001 | 0.31 | <0.001 |
muscles | 19.83 ± 2.93 | 85.11 ± 17.51 | 16.52 ± 3.41 | 630.74 | <0.001 | <0.001 | 0.23 | <0.001 |
lung field | 23.17 ± 4.68 | 66.64 ± 15.79 | 23.00 ± 4.27 | 252.17 | <0.001 | <0.001 | 0.99 | <0.001 |
fats | 19.48 ± 3.96 | 69.26 ± 14.68 | 15.70 ± 3.10 | 547.73 | <0.001 | <0.001 | <0.001 | <0.001 |