Table 1 NEMA Measurements performed in this study and its main differences.

From: NEMA Performance Evaluation of CareMiBrain dedicated brain PET and Comparison with the whole-body and dedicated brain PET systems

Measurement

Phantom/Reconstruction Details

Differences

NU 4-2008

NU 2-2007

NU 2-2012

Spatial Resolution

Encapsulated 22Na Source. Reconstructed with FBP

Capillary filled with 18F. Reconstructed with FBP

Capillary filled with 18F. Reconstructed with FBP

- Different positions for all three protocols

- 2012 admits alternative iterative algorithms but is mandatory FBP reporting

Scatter fraction, count losses, random coincidences measurements

Cylindrical polyethylene phantom. Different dimensions for mouse, rat and monkey

Cylindrical polyethylene phantom for whole body.

Cylindrical polyethylene phantom for whole body.

- Different phantom in 2008

- Different phantom positioning and different tolerances for the activity filling between 2007–2012

Sensitivity

Encapsulated 22Na Source

Capillary filled with 18F and different sleeves sizes phantom

Capillary filled with 18F and different sleeves sizes phantom

- Different phantom in 2008

- Slightly different parameter calculation and tolerances for the activity filling between 2007–2012.

Image Quality, accuracy of attenuation correction and scatter correction

Customized phantom that includes uniformity region, rods and cold rods (air/water). Filled with 18F

Customized phantom that includes: uniformity region, lung insert, and 6 spheres (hot regions). Filled with 18F

Customized phantom that includes: uniformity region, lung insert, and 6 spheres (hot regions). Filled with 18F

- Different phantom in 2008

- Different phantom positioning, acquisition time and different tolerances for the activity filling between 2007–2012

Accuracy: corrections for count losses and randoms

Not Included

The same phantom as the scatter fraction test

The same phantom as scatter fraction test

- Different phantom positioning and different tolerances for the activity filling between 2007–2012

- The reconstructed image values are compared to different references between 2007–2012