Table 8 PET image quality (contrast recovery) for different sphere diameters (whole-body PETs).

From: NEMA Performance Evaluation of CareMiBrain dedicated brain PET and Comparison with the whole-body and dedicated brain PET systems

PET Name

Contrast Recovery (Background Variability)

10 mm

13 mm

17 mm

22 mm

28 mm

37 mm

Relative count rate error at NECR-peak (%)

Celestion

27,9 (6,2)

48,6 (5,3)

52,0 (4,6)

60,5 (4,1)

72,9 (3,9)

77,1 (3,8)

17.3 (TOF-LM-OSEM)

Biograph mCT flow

28,5 (24,3)

42,3 (42,6)

58,4 (568,35)

71,7 (70,7)

70,1 (70,3)

78,3 (78,25)

3.7 (FORE-FBP)

Biograph mCT

37

66

72,2

77,3

81

85,3

1.9 (FORE-FBP)

Biograph mMR

23,4

57,1

75,2

83

58,8

67,5

4.9 (FPRE-FBP)

Vereos

38,4 (8.4)

61,3 (7.1)

65,9 (5,9)

68.5 (4,6)

83,4 (3,5)

86,4 (2,8)

Ingenuity TF

0,39

0,73

0,74

0,8

0,84

0,81

Ingenuity PET/MR

30 (7)

50 (6)

61 (4)

70 (3)

78 (3)

81 (3)

Geminity

34,70 (8,65)

51,43 (7.89)

60,83 (7.05)

72,13 (6.15)

73.88 (5.46)

77.65 (5.04)

 

SIGNA PET/MR

35,2 (4,9)

48,9 (4,0)

59,9 (3,2)

68,6 (2.7)

79,2 (2.2)

87.4 (1.9)

3.3 (OSEM)

Discovery MI

51.7 (10)

61.5 (7.8)

66.2 (6.0)

81.3 (4.8)

86.6 (3.8)

90.0 (3.0)

2–3 (not in NECR-peak)

Discovery IQ

22 (3.3)

68 (2.0)

68 (2.0)

76 (2.5)

73 (2.5)

81 (2.5)

3.9 (OSEM)