Table 1 Wedge testing.
From: Wedge prism approach for simultaneous multichannel microscopy
α (deg) | θd (deg) | objective | Δexp (mm) | Δtheor (mm) | % diff | Twedge/Tdual-view |
|---|---|---|---|---|---|---|
1.93 | 1.00 | 10x | 3.55 | 3.49 | 1.7% | nt |
1.93 | 1.00 | 20x | 3.54 | 3.49 | 1.3% | nt |
1.93 | 1.00 | 40x | 3.55 | 3.49 | 1.6% | nt |
1.93 | 1.00 | 60x | 3.54 | 3.49 | 1.3% | nt |
1.93 | 1.00 | 100x | 3.54 | 3.49 | 1.5% | nt |
1.24 | 0.64 | 10x | 2.31 | 2.23 | 3.6% | 0.58 ± 0.07 |
1.24 | 0.64 | 20x | 2.32 | 2.23 | 3.7% | 0.65 ± 0.10 |
1.24 | 0.64 | 40x | 2.32 | 2.23 | 3.9% | 0.96 ± 0.08 |
1.24 | 0.64 | 60x | 2.32 | 2.23 | 3.9% | 0.57 ± 0.06 |
0.97 | 0.50 | 60x | 1.74 | 1.75 | −0.1% | nt |