Table 1 Conductivity of various KCl solutions determined by local electric field measurements.

From: Development of glass micro-electrodes for local electric field, electrical conductivity, and pH measurements

c0 [mol/L]

Conductivity [S/m]

Difference [%]

Present

Reference37,38

5.0 × 10−4

7.93 × 10−3

7.3905 × 10−3

7.3

1.0 × 10−3

1.67 × 10−2

1.4695 × 10−2

13.6

5.0 × 10−3

6.92 × 10−2

7.1775 × 10−2

3.6

1.0 × 10−2

1.46 × 10−1

1.4127 × 10−1

3.5