Table 4 Time effort evaluation of craniospinal axis irradiation for ten patients treated using the Hi-ART tomotherapy system. The last line indicates mean values over the ten treatment cases. MU stands for monitor units. MF stands for modulation factor.

From: Helical tomotherapy: Comparison of Hi-ART and Radixact clinical patient treatments at the Technical University of Munich

Diagnosis

Irradiation time [s]

Number of gantry rotations

Gantry period [s]

Expected MU

Couch travel [cm]

Couch speed [mm/s]

MF

Field Width [cm]

Pitch

fractions

Leukemia

788.4

49.3

16

11260

71

0.9

3.0

5

0.287

16

Astrocytoma

695.9

35.7

19.5

9918

77

1.11

2.0

5

0.43

20

Ependymoma

774.1

51.6

15

11050

74.4

0.96

2.5

5

0.287

20

Medulloblastoma

1118.6

79.9

14

16036

79.9

0.71

1.8

2.5

0.4

22

Medulloblastoma

933.4

59.5

15.7

13356

85.7

0.92

2.4

5

0.287

20

Leptomeningeal spread

771.8

60.3

12.8

11020

86.9

1.12

2.2

5

0.287

20

Leptomeningeal spread

1023.8

53.9

19

14665

77.6

0.76

2.9.

5

0.287

24

Mesencephalic germinoma

1620.9

108.1

15

23300

77.5

0.48

2.7

2.5

0.287

15

Multiple metastases of the central nervous system

736

26.8

27.5

10500

57.8

0.78

2.5.

5

0.43

20

Multiple metastases of the central nervous system

696.2

43.5

16

9907

74.3

1.07

2.1

5

0.34

20

Mean value (Median value for field width)

915.9

56.9

17.1

13101.2

76.2

0.88

2.4

45

0.33

20