Figure 6
From: Magnetic resonance fingerprinting of the pancreas at 1.5 T and 3.0 T

T1 and T2 phantom values at both 1.5 T and 3.0 T, using conventional techniques and MRF. T1 values between MRF and FSE IR had low absolute mean differences (< 0.05 s for all vials above 0.1 s). There were four vials where the 2D VFA, 3D VFA, and 2D MRF all failed, which occurred at the centre of the T2-layer of the NIST phantom where the T2 values were < 30 ms. The MSE flattened at longer T2 values. MOLLI-T1 and 3D VFA was higher than both MRF-T1 and FSE-IR-T1.