Table 1 Individual surgery date, electrode and estimated position by observer 1 and 2 for DVT/CT and MRI.
From: Evaluation of cochlear implant electrode scalar position by 3 Tesla magnet resonance imaging
No | Surgery | Electrode | Observer 1 position DVT/CT | Observer 2 position DVT/CT | Observer 1 and 2 position MRI |
|---|---|---|---|---|---|
20 | 5.3.20 | Slim J | ST | ST | ST |
19 | 4.5.20 | Flex 28 | ST | ST | ST |
18 | 27.5.20 | Flex 28 | ST | ST | ST |
17 | 13.8.20 | Flex 28 | ST | ST | ST |
16 | 18.8.20 | Flex 28 | ST | ST | ST |
15 | 22.9.20 | Flex 28 | ST | ST | ST |
14 | 13.10.20 | Flex 28 | ST | ST | ST |
13 | 15.10.20 | HFMS | ST | ST | ST |
12 | 29.10.20 | HFMS | ST | ST | ST |
11 | 6.11.20 | HFMS | ST | ST | ST |
10 | 15.11.20 | HFMS | Translocation | ST | Translocation |
9 | 17.11.20 | Slim J | ST | ST | ST |
8 | 26.11.20 | HFMS | ST | ST | ST |
7 | 1.12.20 | Flex 28 | ST | ST | ST |
6 | 3.12.20 | Flex 28 | ST | ST | ST |
5 | 15.12.20 | Flex 28 | ST | ST | ST |
4 | 16.12.20 | Slim J | ST | ST | ST |
3 | 18.12.20 | Slim J | ST | ST | ST |
2 | 8.1.21 | Flex 28 | ST | ST | ST |
1 | 28.1.21 | Slim J | ST | ST | ST |