Table 3 Compensation errors in \({\mathscr {T}}_{\mathrm{val}}\) (diameter 0.6 m), for simulations with mirroring (\({\text {n}} = 7\)) and without mirroring (\({\text {n}} = 0\)).

From: Magnetic field compensation coil design for magnetoencephalography

Coil

Error, \(RDM/\%\) level \(n = 7\)

Error, \(MRD/\%\) level \(n = 7\)

Error, \(RDM/\%\) level \(n = 0\)

Error, \(MRD/\%\) level \(n = 0\)

\(C_{x,\,{\mathrm{hom}}}\)

0.07

0.30

0.55

1.78

\(C_{y,\,{\mathrm{hom}}}\)

0.06

0.23

3.40

10.31

\(C_{z,\,{\mathrm{hom}}}\)

0.12

0.43

1.60

6.51

\(C_{x,\,\nabla _y}\)

0.48

0.88

17.30

18.84

\(C_{x,\,\nabla _x}\)

0.10

0.14

22.32

15.36

\(C_{z,\,\nabla _z}\)

0.17

0.44

15.56

11.89

\(C_{x,\,\nabla _z}\)

0.03

0.05

0.59

1.71

\(C_{z,\,\nabla _y}\)

0.77

1.10

27.30

31.06