Table 2 Coefficients of the best fit curves, with 95% confidence interval (round brackets), to the normalized dose profile as a function of the z distance and the t thickness.

From: Modelling a new approach for radio-ablation after resection of breast ductal carcinoma in-situ based on the BAT-90 medical device

t (mm)

z0 (mm)

a1 (Gy cm2/MBq)

a2 (mm−1)

0

0

170.9 (166.1, 175.7)

− 7.2 (− 7.6, 6.8)

1

0.5

115.7 (112.5, 118.9)

− 6.7 (− 6.9, -6.4)

3

1.5

68.5 (66.3, 70.8)

− 6.7 (− 7.0, -6.4)

5

2.5

46.3 (46.3, 47.83)

− 6.8 (− 7.11, -6.5)