Table 2 Algorithm sensitivity results.

From: Automated detection of cerebral microbleeds on T2*-weighted MRI

 

Definite

Probable

Combined

SWI

Rater identified

39

15

54

Algorithm identified

38

12

50

Sensitivity

0.97

0.8

0.93

GRE

Rater identified

43

18

61

Algorithm identified

41

15

56

Sensitivity

0.95

0.83

0.92

Merged

Rater identified

45

19

64

Algorithm identified

44

17

61

Sensitivity

0.98

0.89

0.95

  1. An artifact was labeled as a “definite” microbleed if all three raters agreed that the artifact was a microbleed, and as a “probable” microbleed if two raters agreed that it was a microbleed. This table shows the sensitivity results of the algorithm across these different labels. Note that in a study using only visual ratings, the final column (combining the definite and probable ratings) would be the number typically reported.