Table 1 Minimal lumen diameter of stented coronary artery segments measured by quantitative coronary angiographic (QCA).
FIREHAWK | JDBM BRS | P value | |
|---|---|---|---|
Pre-procedure, n | 4 | 4 | |
Reference vessel diameter, mm | 2.63 ± 0.04 | 2.61 ± 0.05 | 0.6523 |
Minimal luminal diameter, mm | 2.60 ± 0.08 | 2.56 ± 0.07 | 0.6443 |
30 day follow-up, n | 4 | 4 | |
Minimal lumen diameter, mm | 2.49 ± 0.13 | 2.40 ± 0.15 | 0.4577 |
90 day follow-up, n | 4 | 4 | |
Minimal lumen diameter, mm | 2.33 ± 0.15 | 2.18 ± 0.23 | 0.3455 |
180 day follow-up, n | 4 | 4 | |
Minimal lumen diameter, mm | 2.13 ± 0.11 | 2.01 ± 0.25 | 0.2165 |