Table 1 Minimal lumen diameter of stented coronary artery segments measured by quantitative coronary angiographic (QCA).

From: Biosafety and efficacy evaluation of a biodegradable magnesium-based drug-eluting stent in porcine coronary artery

 

FIREHAWK

JDBM BRS

P value

Pre-procedure, n

4

4

 

Reference vessel diameter, mm

2.63 ± 0.04

2.61 ± 0.05

0.6523

Minimal luminal diameter, mm

2.60 ± 0.08

2.56 ± 0.07

0.6443

30 day follow-up, n

4

4

 

Minimal lumen diameter, mm

2.49 ± 0.13

2.40 ± 0.15

0.4577

90 day follow-up, n

4

4

 

Minimal lumen diameter, mm

2.33 ± 0.15

2.18 ± 0.23

0.3455

180 day follow-up, n

4

4

 

Minimal lumen diameter, mm

2.13 ± 0.11

2.01 ± 0.25

0.2165