Table 3 Differences in SpO2 measurements between pulse oximeters.
Mean ± SD (95% CI) | P-value | |
|---|---|---|
SaO2 < 94% (Blood gas analysis = 274) | ||
Nihon Kohden | 1.41 ± 2.28% (1.14–1.68) | 0.083 |
Masimo | 1.74 ± 2.15% (1.49–2.00) | |
94% ≤ SaO2 < 98% (Blood gas analysis = 952) | ||
Nihon Kohden | 0.57 ± 1.77% (0.46–0.69) | < 0.0001 |
Masimo | 1.10 ± 1.60% (1.00–1.20) | |
SaO2 ≥ 98% (Blood gas analysis = 271) | ||
Nihon Kohden | 0.51 ± 1.49% (0.33–0.69) | 0.006 |
Masimo | 0.84 ± 1.28% (0.68–0.99) | |