Table 1 Repeatability outcomes for biometric measurements obtained using the ANTERION, CASIA II and IOLMaster500.
Parameters | Devices | Mean (± SD) | RC (95% CI) | p-value for pairwise comparison in RC | ||
|---|---|---|---|---|---|---|
ANTERION vs CASIAII | ANTERION vs IOLMaster500 | CASIAII vs IOLMaster500 | ||||
Km (D) | ANTERION | 43.878 ± 1.415 | 0.2345 (0.2010–0.2681) | 0.600 | 0.138 | 0.254 |
CASIA II | 43.851 ± 1.440 | 0.2484 (0.2125–0.2843) | ||||
IOLMaster500 | 43.902 ± 1.456 | 0.3148 (0.2666–0.3629) | ||||
J0 (D) | ANTERION | 0.305 ± 0.457 | 0.2003 (0.1716–0.2289) | 0.002 | 0.039 | 0.361 |
CASIA II | 0.376 ± 0.422 | 0.3217 (0.2752–0.3682) | ||||
IOLMaster500 | 0.207 ± 0.469 | 0.5448 (0.4614–0.6282) | ||||
J45 (D) | ANTERION | − 0.002 ± 0.180 | 0.2232 (0.1913–0.2551) | 0.980 | 0.245 | 0.072 |
CASIA II | 0.029 ± 0.196 | 0.2115 (0.1809–0.2420) | ||||
IOLMaster500 | − 0.024 ± 0.226 | 0.3336 (0.2825–0.3846) | ||||
ACD (mm) | ANTERION | 2.899 ± 0.597 | 0.0311 (0.0264–0.0352) | 0.216 | < 0.001 | < 0.001 |
CASIA II | 2.912 ± 0.586 | 0.0260 (0.0227–0.0298) | ||||
IOLMaster500 | 2.932 ± 0.563 | 0.1947 (0.1652–0.2241) | ||||
CCT (µm) | ANTERION | 543.17 ± 32.20 | 2.1759 (1.8648–2.4869) | 0.575 | – | – |
CASIA II | 543.02 ± 33.45 | 2.3223 (1.9868–2.6579) | ||||
AL (mm) | ANTERION | 23.683 ± 1.137 | 0.0154 (0.0130–0.0178) | – | 0.002 | – |
IOLMaster500 | 23.881 ± 1.323 | 0.0260 (0.0221–0.0299) | ||||