Table 1 Performance characteristics of enzymatically calibrated PAM-AMA Assay.

From: Novel insights into peptide amidation and amidating activity in the human circulation

Analytical performance characteristics of the AMA assay

Calibrator range

0.712–121.865 AMA units (0–4000 ng/mL)

Intra-assay CV

3–6%

Limit of detection

1.94 AMA units (42.4 ng/mL)

Limit of quantification

2.44 AMA units (60.4 ng/mL)

Accuracy (recovery of added PAM to serum)

Average (range) % recovery from expected = 108% (103–116%)

Linearity (mixing studies) 12 ×

Range: 3.2–16.5 AMA units

Average (range) % recovery from expected = 91% (82–101.6%)

Dilution linearity (10–100 × in assay)

Range: 0–16.5 AMA units

Average (range) % recovery from expected = 98% (91–103%)

Variation of substrate concentration CV

Range: 10–40 ng/mL; CV: 1.2–2.4%